Overview As a professional sterilization manufacturer which has the
production experience over ten years, we believe that our advanced
technology will widen the choice for your experiments. Work process of sterilization and drying a: sterilization temperature decreases to 102℃, then exhaust and
drain b: sterilization, exhaust and drain after sterilization c: sterilization, exhaust, drain, drying Technical Parameter Model: YXQ-LS-50G Volume: 50L Max. working temperature: 135℃ Max. working pressure: 0.22MPa Power Consumption of sterilization: 3.1KW Power Consumption of drying: 0.8KW Power supply: 220V±10%/50Hz±2% DIM. Exterior (mm): 600*500*1080 DIM. Interior (mm): Ф350*525 DIM. Basket (mm): Ф330*250*2 Features Auto-control type with drying function Structure of hand wheel type quick-door-open safety interlock
device (accord with the 49th requirement of Vessel Regulations) Material: outer and chamber are made of high quality stainless
steel SUS304, acid proof, alkali resistance, anti-corrosive Microcomputer intelligent control Pressure safety interlock device, over temperature protection
device Auto-raising type packing ring, exhaust cold air automatically Auto-exhause when over pressure, low water level alarm,
auto-control when lack of water Temperature range (109℃-135℃), sterilization time range (4-120min),
drying time range (drying after sterilization) (30-240min), Max.
design/work pressure (0.25MPa-0.22MPa) Inner circulation exhaust type, with 3L steam collection bottle Automatic inlet, heating, sterilization, exhaust and drying Buzzer reminds after work and close-down automatically